Flow rate = [(VOUT-1 V) / 4 V] Full range flow rate
For example: CAFS 5019AV has 1000 slpm a full range flow. When the output voltage is 2.5V to read, the instantaneous flow rate is
[(2.5V-1V)/4V × 1000slpm ] =375 slpm
Flow rate = [(VOUT-1 V) / 4 V] Full range flow rate

|
Functional items | CMF5019 Flow Range 0-60 M/S Medical Flow Sensors | ||
| performance parameter | unit | remarks | |
| Power Supply Voltage | 8~24VDC ,50mA | VDC | selectable |
| accuracy | ±1.5 | % | FS |
| response time | 65 | ms | selectable |
| maximum pressure | 0.8 | MPa | customizable |
| communication mode | RS 485(MODBUS ) | ||
|
Output method | Analog outputs of 1-5 | VDC | |
| 4-20mA | apolegamy | ||
| temperature | Medium temperature (-10~65), ambient temperature (-25~85) | ℃ | |
| joggle | G3/4 | customizable | |
| Calibration method | Air, and the 20℃,101.325kPa | ||
| The overall material | stainless steel | ||